top of page
Safety drying chamber

Safety drying chamber

The Safety Drying Chamber is a crucial piece of equipment for any medical facility. With a silicone and dust-free inner chamber, this drying chamber is designed to ensure the safe and efficient drying of specimens containing solvents. The symmetric airflow within the chamber guarantees even and thorough drying, providing reliable results every time. This safety drying oven offers peace of mind for lab technicians and researchers, knowing that their delicate and potentially hazardous specimens are being handled with the utmost care. With its advanced features and safety measures, this drying chamber is an essential tool for any medical setting.
  • Benefits

    • Safety concept meets the DIN EN 1539-2010-08 standard
    • Short drying time with high temperature accuracy and defined ventilation
    • Silicone free and dust-free inner chamber
    • Replaceable fresh air cartridges are easy to change and ensure constant fresh air quality
  • IMPORTANT CHARACTERISTICS

    • Temperature range: +10°C above ambient temperature to +300 °C
    • 2 chrome-plated racks
    • Controller with time-segment and real-time programming
    • Fresh-air monitoring with audible and visual alarm and automatic deactivation of heating
    • Replaceable filter cartridge, Class M6
    • Class 2 independent adjustable temperature safety device with visual alarm
    • Standard door gasket made from solvent-resistant FKM for use at up to 200 °C
    • Computer interface: RS 422

    • Optional door gasket made from silicone with limited solvent resistance for use at up to 300 °C

  • Performance Data Temperature

    Temperature range +10 °C above ambient temperature to 300 °C
    Temperature uniformity at 150°C 2.5 ± K
    Temperature fluctuation at 150°C 0.8 ± K
    Heating up time to 150°C 20 min
    Recovery time after door was opened for 30 s at 150°C 12 min
  • Technical specifications

    Performance Data Temperature

    Temperature range +10 °C above ambient temperature to 300 °C
    Temperature uniformity at 150°C 2.5 ± K
    Temperature fluctuation at 150°C 0.8 ± K
    Heating up time to 150°C 20 min
    Recovery time after door was opened for 30 s at 150°C 12 min

     

    Air change data

    Air circulation (approx.) 20 x/min
    Volumetric flow rate of exhaust air acc. to EN 1539 at 50 °C 400 ca. L/min

     

    Solvent quantity

    Highest permitted solvent quantity (at T-180°C, M-100g/mol, U-40g/m3, K=0,5) 4,67g

     

    Electrical data

     

    Rated Voltage 230 V
    Power frequency 50/60 Hz
    Nominal power 2,9 kW
    Phase (Nominal voltage) 1~

Related Products