top of page
Elisa Reader

Elisa Reader

The Elisa Reader is a state-of-the-art medical device designed to accurately and efficiently measure the results of Enzyme-Linked Immunosorbent Assays (ELISA) in a clinical or research setting. This innovative equipment features a user-friendly interface, high sensitivity, and precise detection capabilities, ensuring reliable and reproducible test results. With its automated scanning and analysis features, the Elisa Reader streamlines the laboratory workflow, saving time and reducing the risk of human error. Its compact design and robust construction make it a dependable tool for any laboratory environment. Whether used for diagnosing diseases, monitoring treatment outcomes, or conducting research experiments, the Elisa Reader is a essential addition to any modern medical facility.
  • Applications

    • Medical Diagnostics: ELISA readers are extensively used in medical diagnostics to detect various diseases and infections. They can detect specific antibodies or antigens present in patient samples such as blood, serum, or urine. For example, ELISA readers are commonly used in the diagnosis of HIV, hepatitis, Lyme disease, and various autoimmune disorders.
    • Drug Development and Quality Control: ELISA readers are vital tools in pharmaceutical research and development. They are used to quantify the presence of specific proteins or biomarkers in cell cultures or biological samples. ELISA assays can be employed to assess drug efficacy, toxicity, and pharmacokinetics during the drug development process. They are also utilized for quality control to ensure the consistency and purity of pharmaceutical products.
    • Food Safety and Environmental Monitoring: ELISA readers play a crucial role in monitoring food safety and environmental contamination. They can detect the presence of allergens, toxins, pathogens, and contaminants in food and environmental samples. ELISA assays are commonly used in industries such as agriculture, food processing, and water treatment to ensure compliance with regulatory standards and to safeguard public health.
    • Research in Life Sciences: ELISA readers are extensively used in basic and clinical research across various fields of life sciences. Researchers use ELISA assays to study biomolecular interactions, quantify protein expression levels, and investigate disease mechanisms. ELISA readers enable researchers to perform high-throughput screening of samples and generate quantitative data for statistical analysis.
    • Veterinary Diagnostics: ELISA readers are employed in veterinary medicine for the diagnosis of infectious diseases in animals. They can detect specific antibodies or antigens associated with diseases such as canine parvovirus, feline leukemia virus, and bovine viral diarrhea virus. ELISA assays are valuable tools for monitoring and controlling disease outbreaks in livestock populations and companion animals.
    • Biodefense and Biosecurity: ELISA readers are used in biodefense and biosecurity applications to detect the presence of biothreat agents such as anthrax, botulinum toxin, and ricin. They are employed in laboratories and field settings for rapid and sensitive detection of biological warfare agents and emerging infectious diseases. ELISA assays are integral components of bio-surveillance systems aimed at early detection and response to biosecurity threats.
  • TECHNICAL DETAILS

    Dispenser

    Type 4 independent precision pumps
    Volume Range 10-2000 µL
    Precision <2% CV at 100 µL
    Accuracy <5% at 100 µL
    Reagent support Any bottle or vial type can be attached

     

    Washer

    Manifold configuration

    Autoclavable 8-way manifold with 2 parallel needles

    Dispense volume 25-1000 µL
    Wash modes Standard, Overflow, Soak, Two-Point, Sweep
    Wash buffer capacity 3 x 1 L bottles for different wash solutions
    Waste liquid container 2 L bottle

     

    Incubator

    Temperature range From ambient +4° C to 55° C
    Temperature stability ±1° C across plate at 37° C
    Incubation time Programmable
    Temperature monitoring Yes

     

    Shaker

    Shaking Independent linear motion 5-20 Hz

     

    Software

    Software  
    • Crocodile Control Software for protocol creation and running
    • Optional: LightCompass for data analysis, easy automation and compliance with CFR 21 part 11

     

    General specifications

    Sample format 1 x 96-well microplate (ANSI SLAS)
    Power supply Desktop power supply and power cord. Input 110-240 V AC, output 24 V/3.3 A DC
    Dimensions (W x D x H) 263 x 626 x 258 mm
    Weight 14.8 kg
    Interface USB
    Operating temperature 10-35° C
    Humidity Up to 75% (non condensing)

     